ADMET Online Troubleshooting
EP Digital Controller

DATA ACQUISITION/ANALYSIS PROBLEMS

Problem Description

Possible Causes (click entries for applicable details)

Analysis Result is 0 or N/A

Parameter specification for data range that is not being collected during test. Many analyses have user selectable parameters such as displacement or Samplebreak %, which is the percentage load dropfrom maximum load. These settings need to be consistent with the data being logged

Test Ending Early

Sample Break % and/or Data Logging Threshold (Data Acq menu) not set to values appropriate to your test

Local Overload (Data Acq menu) set to low

Log Rate (Data Acq menu) not set to appropriate value

Testing Machine Movement Problems

Machine not moving at requested rate

Tuning Gains (Control menu) not set properly

Incorrect Test Method is active. NOTE: tuning gains which control machine movement are test method specific

Machine is not moving

Tuning Gains (Control menu) not set properly

Incorrect Test Method is active. NOTE: tuning gains which control machine movement are test method specific

Postion Calibration value is incorrect

Machine moves in wrong direction during testing

Direction of movement specified incorrectly. Control menu -> Dir

Machine moves in wrong direction during homing

Home Location is not set properly

 