ADMET Online Troubleshooting
Gauge Buster 2 Digital Indicator
| Problem Description - Not Holding Peak Load |
| Possible Cause - Sample Break threshold and sample break % settings |
Action - NOTE: The Gauge Buster 2 digital goes into TESTING
MODE when the actual load exceeds the programmed load
threshold value. When the Gauge Buster 2 Digital is in the
TESTING MODE, the live load is displayed on the upper
right and the load rate bar or load rate is displayed in the
bottom of the display. If you do not see the load rate bar or
the numeric load rate in the bottom display then you are NOT
in the testing mode and you need to look at your data logging
threshold value.
Ensure that the Break Threshold (Sample Brk menu item in
the Setup menu) is set to a value that is appropriate for the
test. Break Threshold is the point where the Gauge Buster 2
digital is armed for sample break. If the break threshold is too
low then a load drop in the early part of the test could trigger
the Gauge Buster 2 to detect premature specimen break (end
of test) and stop logging data. For example, if the Break
Threshold is set to 10lbs (much too low) and the sample break
percentage is set to 50% and the load climbed to 20lbs and
then dropped to 10lbs the Gauge Buster 2 would detect
sample break, stop logging data and report a peak load of
20lbs.
Ensure that Sample Break (SETUP, End of Test menu) is set
appropriately. Sample Break is defined as a percentage of
peak load and is used to define the end of a test. If Sample
Break = 10 percent and Peak Load = 100,000 Lb, then the
test will terminate when the load drops below 10,000 Lb.
Setting sample break percentage to 0 disables it so the Gauge
Buster 2 will continue logging data until the test buffer is full. If
the sample break percentage is set too high then end of test
can be triggered early.
|
| Problem Description - Not Holding Peak Load |
| Possible Cause - Applying Load too quickly after taring Gauge Buster PLUS |
Action - NOTE: The Gauge Buster 2 is designed for hands-free operation. You should only need to zero the digital once. It is not necessary and NOT recommended to zero the digital prior to each test. Doing so can prevent the digital from going into
Testing Mode and not recording the peak load. If you do not see the rate bar or load rate display on bottom line of the
display it then the digital is not going into testing mode.
Ensure that you are that you are waiting at least 3 seconds after zeroing digital before running a test (applying load to your
break machine).
|
| Problem Description - Not tracking load while machine actually loading and breaking sample |
| Possible Cause - Sample Break threshold and sample break % settings |
Action - NOTE: The Gauge Buster 2 digital goes into TESTING
MODE when the actual load exceeds the programmed load
threshold value. When the Gauge Buster 2 Digital is in the
TESTING MODE, the live load is displayed on the upper
right and the load rate bar or load rate is displayed in the
bottom of the display. If you do not see the load rate bar or the numeric load rate in the bottom display then you are NOT in the testing mode and you need to look at your data logging
threshold value.
Ensure that the Break Threshold (Sample Brk menu item in the Setup menu) is set to a value that is appropriate for the test. Break Threshold is the point where the Gauge Buster 2 digital is armed for sample break. If the break threshold is too low then a load drop in the early part of the test could trigger the Gauge Buster 2 to detect premature specimen break (end of test) and stop logging data. For example, if the Break Threshold is set to 10lbs (much too low) and the sample break percentage is set to 50% and the load climbed to 20lbs and then dropped to 10lbs the Gauge Buster 2 would detect sample break, stop logging data and report a peak load of 20lbs.
Ensure that Sample Break (SETUP, End of Test menu) is set appropriately. Sample Break is defined as a percentage of peak load and is used to define the end of a test. If Sample Break = 10 percent and Peak Load = 100,000 Lb, then the test will terminate when the load drops below 10,000 Lb. Setting sample break percentage to 0 disables it so the Gauge Buster 2 will continue logging data until the test buffer is full. If the sample break percentage is set too high then end of test can be triggered early.
|
| Problem Description - Not tracking load while machine actually loading and breaking sample |
| Possible Cause - Applying Load too quickly after taring Gauge Buster PLUS |
Action - NOTE: The Gauge Buster 2 is designed for hands-free operation. You should only need to zero the digital once. It is not necessary and NOT recommended to zero the digital prior to each test. Doing so can prevent the digital from going into Testing Mode and not recording the peak load. If you do not see the rate bar or load rate display on bottom line of the display it then the digital is not going into testing mode.
Ensure that you are that you are waiting at least 3 seconds after zeroing digital before running a test (applying load to your break machine).
|
| Problem Description - Rate bar not being displayed |
| Possible Cause - Sample Break threshold and sample break % settings |
Action - NOTE: The Gauge Buster 2 digital goes into TESTING
MODE when the actual load exceeds the programmed load
threshold value. When the Gauge Buster 2 Digital is in the
TESTING MODE, the live load is displayed on the upper
right and the load rate bar or load rate is displayed in the
bottom of the display. If you do not see the load rate bar or the numeric load rate in the bottom display then you are NOT in the testing mode and you need to look at your data logging
threshold value.
Ensure that the Break Threshold (Sample Brk menu item in the Setup menu) is set to a value that is appropriate for the test. Break Threshold is the point where the Gauge Buster 2 digital is armed for sample break. If the break threshold is too low then a load drop in the early part of the test could trigger the Gauge Buster 2 to detect premature specimen break (end of test) and stop logging data. For example, if the Break Threshold is set to 10lbs (much too low) and the sample break percentage is set to 50% and the load climbed to 20lbs and then dropped to 10lbs the Gauge Buster 2 would detect sample break, stop logging data and report a peak load of 20lbs.
Ensure that Sample Break (SETUP, End of Test menu) is set appropriately. Sample Break is defined as a percentage of peak load and is used to define the end of a test. If Sample Break = 10 percent and Peak Load = 100,000 Lb, then the test will terminate when the load drops below 10,000 Lb. Setting sample break percentage to 0 disables it so the Gauge Buster 2 will continue logging data until the test buffer is full. If the sample break percentage is set too high then end of test can be triggered early.
|
| Problem Description - Rate bar not being displayed |
| Possible Cause - Applying Load too quickly after taring Gauge Buster PLUS |
Action - NOTE: The Gauge Buster 2 is designed for hands-free operation. You should only need to zero the digital once. It is not necessary and NOT recommended to zero the digital prior to each test. Doing so can prevent the digital from going into Testing Mode and not recording the peak load. If you do not see the rate bar or load rate display on bottom line of the display it then the digital is not going into testing mode.
Ensure that you are that you are waiting at least 3 seconds after zeroing digital before running a test (applying load to your break machine).
|
| Problem Description - Load reading wrong |
| Possible Cause - Calibration |
Action - Ensure that correct load calibration is selected. The Active Calibration # is displayed in the upper left of the live screen. The Gauge Buster 2 can store up to six load cell calibrations.
|
| Problem Description - Load reading wrong |
| Possible Cause - Analog/Digital Electronics |
Action - Contact ADMET technical support.
|
| Problem Description - Load reading wrong |
| Possible Cause - Excitation Voltage bad |
Action - Contact ADMET technical support.
|
| Problem Description - Load reading wrong |
| Possible Cause - Load Transducer Cable |
Action - Replace/fix transducer cable.
NOTE: refer to See Electrical/Mechanical Data section of this manual for cable wiring information.
|
| Problem Description - Load reading wrong |
| Possible Cause - Load Transducer |
Action - Replace Load Transducer.
|
| Problem Description - Stress reading wrong |
| Possible Cause - Specimen Information |
Action - Stress is a calculated value. It is equal to the load divided by the programmed specimen cross-sectional area.
Ensure that Specimen Type and specimen dimension information is correct in the SETUP menu
|
| Problem Description - Load Reading Unstable |
| Possible Cause - Calibration |
Action - Ensure that there is valid calibration in selected load channel. A valid calibration has at least two points, the first point MUST be zero, appropriate full scale and resolution settings, and a good A/D count span between the points in the calibration.
|
| Problem Description - Load Reading Unstable |
| Possible Cause - Analog/Digital Electronics |
Action - Contact ADMET technical support.
|
| Problem Description - Load Reading Unstable |
| Possible Cause - Excitation Voltage bad |
Action - Contact ADMET technical support.
|
| Problem Description - Load Reading Unstable |
| Possible Cause - Load Transducer Cable |
Action - Ensure that cable gain strap is properly wired for transducer being used.
Replace/fix transducer cable.
NOTE: refer to See Electrical/Mechanical Data section of this manual for cable wiring information.
|
|
Contact sales@admet.com or
visit http://www.admet.com/formNotSure.htm to see how ADMET can help and to
find your local ADMET distributor.