ISO 14801 | Medical Devices and Equipment | Fatigue Testing
ISO 14801 centers on fatigue testing of dental implants and their prosthetic elements. The test has been used to compare differently designed implants or those of varying size.
Testing should be only performed on a finished specimen. The endosseous dental implant should be assembled according to its intended use. Prior to conducting ISO 14801, it is important to read the entire specification in the relevant publication.
ISO 14801 | Medical Devices and Equipment | Fatigue Testing
- If there is evidence of corrosion fatigue, testing should be carried out in a saline bath at a temperature specified in the standard. Otherwise, testing can be conducted in air at 20C.
- Apply compressive loads sinusoidally to the implant until failure as described in the standard. Perform tests in saline to 2 million cycles. Perform tests in air to 5 million cycles.
Calculations:
- Generate a load vs. cycle diagram by showing for each value of applied peak load the number of cycles to failure
- Maximum Load Endured
- Maximum Bending Moment
ISO 14801 | Medical Devices and Equipment | Fatigue Testing
- eXpert 5950 – ISO 14801 in a heated fluid bath
- ISO 14801 Fiture in a Temperature-Controlled Bath
- ISO 14801 on an eXpert 3933 Fatigue Testing Machine
- Close up of sample and fixture for ISO 14801
- MTESTQuattro Test Report for ISO 14801
ISO 14801 | Medical Devices and Equipment | Fatigue Testing
ISO 14801 | Recommended Equipment |
---|---|
Testing System | |
Fixturing |
Ask our Engineers
Just have a quick question about running ISO 14801 on your system?
We're happy to help!
Ask our Engineers